Typ: EP18200E181000Heating block control: with PID controller
Setting the heating block temperature: use the P key and the arrow key to set to 1 ° C
Fine adjustment is automatic
Temperature range: RT ... 200 ° C (at ambient temperature 20 ° C)
Accuracy: ± 1 ° C
Stability: ± 0.1 ° C
Resolution: 1 ° C
Test specimen holder: Ø 18 mm / depth 100 mmDelivery: test certificate power cord sleeve change tool operating manual
Optional accessories: carrying case Transition sleeve DAkkS certificate works certificate The temperature calibrator TP 18200 E consists of two individual housings with a front dimension of 96 x 96 mm, with a depth of 200 mm, which are rigidly connected with each other and provided with a carrying handle at the side. The left housing contains an aluminum heating block with an 18 mm diameter, 100 mm deep, hole accessible from above for receiving the test specimen. The heating cartridge and the temperature sensor for determining the block temperature are installed from below in this heating block.
The heating block is installed insulated in the housing. The right-hand housing contains the complete electronic unit for controlling the heating block temperature. A semiconductor relay (SSR) is used to control the heating cartridges. On the front panel are the controller and a switch for the mains supply. The controller is equipped with a 2-row, 4-digit 7-segment LED for the heating block and setpoint temperature. On the right side of the device combination is the IEC built-in connector for the mains connection. Next to it is the holder for receiving the device control.
The temperature calibrator TP 18200 E consists of two individual housings with a front dimension of 96 x 96 mm, with a depth of 200 mm, which are rigidly connected with each other and provided with a carrying handle at the side. The left housing contains an aluminum heating block with an 18 mm diameter, 100 mm deep, hole accessible from above for receiving the test specimen. The heating cartridge and the temperature sensor for determining the block temperature are installed from below in this heating block.
The heating block is installed insulated in the housing. The right-hand housing contains the complete electronic unit for controlling the heating block temperature. A semiconductor relay (SSR) is used to control the heating cartridges. On the front panel are the controller and a switch for the mains supply. The controller is equipped with a 2-row, 4-digit 7-segment LED for the heating block and setpoint temperature. On the right side of the device combination is the IEC built-in connector for the mains connection. Next to it is the holder for receiving the device control.

Price on request.
delivery time on request
delivery time on request